"""Fluke 8808A 5.5-digit multimeter driver over RS-232.
The 8808A's only remote interface is its RS-232 port, reached as a VISA ASRL resource. It
speaks the Fluke 45-style command set (``VDC``, ``RANGE 3``, ``MEAS1?``), not SCPI, and has no
error queue: the meter reports each command line's outcome with a prompt (``=>`` ok,
``?>`` command error, ``!>`` execution error).
"""
from __future__ import annotations
import pyvisa
from instro.dmm import DMMDriverBase
from instro.dmm.types import MeasurementFunction
from instro.lib import InstroError
from instro.lib.transports.visa import VisaConfig, VisaDriver
_PROMPT_OK = "=>"
_PROMPT_ERRORS = {
"?>": "command error (not understood)",
"!>": "execution error (understood but not executed)",
}
# Standard event status bits QYE | DDE | EXE | CME.
_ESR_ERROR_MASK = 0b0011_1100
_DEVICE_CLEAR = b"\x03"
_PROBE_TIMEOUT_MS = 500
_FUNCTION_COMMAND: dict[MeasurementFunction, str] = {
MeasurementFunction.DC_VOLTAGE: "VDC",
MeasurementFunction.AC_VOLTAGE: "VAC",
MeasurementFunction.DC_CURRENT: "ADC",
MeasurementFunction.AC_CURRENT: "AAC",
# WIRE2/WIRE4 are only accepted once OHMS is active.
MeasurementFunction.TWO_WIRE_RESISTANCE: "OHMS; WIRE2",
MeasurementFunction.FOUR_WIRE_RESISTANCE: "OHMS; WIRE4",
}
_OHMS_RANGES = (200.0, 2e3, 20e3, 200e3, 2e6, 20e6, 100e6)
# Full scale (V, A, ohm) of each ``RANGE <n>`` index, n = 1, 2, ... (manual Table 4-11A).
_RANGES: dict[MeasurementFunction, tuple[float, ...]] = {
MeasurementFunction.DC_VOLTAGE: (0.2, 2.0, 20.0, 200.0, 1000.0),
MeasurementFunction.AC_VOLTAGE: (0.2, 2.0, 20.0, 200.0, 750.0),
MeasurementFunction.DC_CURRENT: (200e-6, 2e-3, 20e-3, 200e-3, 2.0, 10.0),
MeasurementFunction.AC_CURRENT: (20e-3, 200e-3, 2.0, 10.0),
MeasurementFunction.TWO_WIRE_RESISTANCE: _OHMS_RANGES,
MeasurementFunction.FOUR_WIRE_RESISTANCE: _OHMS_RANGES,
}
_DIGITS_RATE = {5: "S", 4: "F"}
[docs]
class Fluke8808A(DMMDriverBase):
"""Fluke 8808A on its RS-232 port, primary display only. Resolution is set via ``set_digits``."""
def __init__(self, visa_resource: str | VisaConfig) -> None:
"""Configure the driver; the port is not opened until ``open()``.
The front panel is locked (``RWLS``) while the driver holds the meter, so the function
it tracks can't be changed underneath it.
Args:
visa_resource (str | VisaConfig): ASRL resource string, e.g. ``"ASRL3::INSTR"`` or
``"ASRL/dev/ttyUSB0::INSTR"``, or a ``VisaConfig`` whose ``serial_config`` matches
the meter's front-panel RS-232 setup. The defaults (9600 baud, 8N1, no flow
control) match the factory setup.
Example:
>>> from instro.dmm import InstroDMM
>>> from instro.unstable.dmm.drivers import Fluke8808A
>>> dmm = InstroDMM(name="fluke", driver=Fluke8808A("ASRL/dev/ttyUSB0::INSTR"))
>>> dmm.open()
>>> dmm.read_dc_voltage()
"""
self._visa = VisaDriver(visa_resource)
self._prompts = True
self._function: MeasurementFunction | None = None
[docs]
def open(self) -> None:
"""Open the port and reset the meter into remote mode with the front panel locked."""
self._visa.open()
try:
with self._visa.lock():
self._device_clear()
self._detect_prompts()
self._transact("*RST")
self._transact("RWLS")
# Bare mantissa/exponent readings, and continuous triggering so MEAS1? returns.
self._transact("FORMAT 1")
self._transact("TRIGGER 1")
self._function = MeasurementFunction.DC_VOLTAGE
except BaseException:
self._visa.close()
raise
[docs]
def close(self) -> None:
"""Return the meter to local control and close the port."""
try:
if self._visa.is_open:
self._transact("LOCS")
finally:
self._visa.close()
self._function = None
[docs]
def set_measurement_function(self, function: MeasurementFunction) -> None:
"""Select ``function`` on the primary display."""
self._transact(_FUNCTION_COMMAND[function])
self._function = function
[docs]
def set_digits(self, n: int) -> None:
"""Set resolution: 5 selects the slow rate (2.5 rdg/s), 4 the fast rate (100 rdg/s).
The medium rate has the same 4.5-digit resolution as fast, so it is not reachable here.
"""
try:
rate = _DIGITS_RATE[n]
except KeyError:
raise ValueError(f"Fluke 8808A supports 4 or 5 digit resolution, got {n}.") from None
self._transact(f"RATE {rate}")
def _set_range(self, function: MeasurementFunction, value: float | None) -> None:
if self._function is not function:
active = self._function.name if self._function else "none"
raise InstroError(
f"Fluke 8808A ranges apply to the active function ({active}); "
f"select {function.name} with set_measurement_function first."
)
if value is None:
self._transact("AUTO")
return
ranges = _RANGES[function]
index = next((i for i, full_scale in enumerate(ranges, start=1) if abs(value) <= full_scale), None)
if index is None:
raise ValueError(f"Fluke 8808A {function.name} range {value:g} exceeds the top range {ranges[-1]:g}.")
self._transact(f"RANGE {index}")
[docs]
def set_dc_voltage_range(self, value: float | None) -> None:
self._set_range(MeasurementFunction.DC_VOLTAGE, value)
[docs]
def set_ac_voltage_range(self, value: float | None) -> None:
self._set_range(MeasurementFunction.AC_VOLTAGE, value)
[docs]
def set_dc_current_range(self, value: float | None) -> None:
self._set_range(MeasurementFunction.DC_CURRENT, value)
[docs]
def set_ac_current_range(self, value: float | None) -> None:
self._set_range(MeasurementFunction.AC_CURRENT, value)
[docs]
def set_two_wire_resistance_range(self, value: float | None) -> None:
self._set_range(MeasurementFunction.TWO_WIRE_RESISTANCE, value)
[docs]
def set_four_wire_resistance_range(self, value: float | None) -> None:
self._set_range(MeasurementFunction.FOUR_WIRE_RESISTANCE, value)
# --- Measurements. MEAS1? waits for the next fresh reading. ---
# An overload ("0L" on the display) reads as +/-1.0E+9.
def _read_value(self, function: MeasurementFunction) -> float:
with self._visa.lock():
# Only re-select on a change: re-selecting the active function may reset its range.
if self._function is not function:
self.set_measurement_function(function)
response = self._transact("MEAS1?")
if response is None:
raise InstroError("Fluke 8808A returned no reading for MEAS1?")
return float(response)
[docs]
def measure_dc_voltage(self) -> float:
return self._read_value(MeasurementFunction.DC_VOLTAGE)
[docs]
def measure_ac_voltage(self) -> float:
return self._read_value(MeasurementFunction.AC_VOLTAGE)
[docs]
def measure_dc_current(self) -> float:
return self._read_value(MeasurementFunction.DC_CURRENT)
[docs]
def measure_ac_current(self) -> float:
return self._read_value(MeasurementFunction.AC_CURRENT)
[docs]
def measure_resistance(self) -> float:
return self._read_value(MeasurementFunction.TWO_WIRE_RESISTANCE)
[docs]
def measure_four_wire_resistance(self) -> float:
return self._read_value(MeasurementFunction.FOUR_WIRE_RESISTANCE)
# --- Line framing ---
def _device_clear(self) -> None:
"""Send ^C to abort any pending command, then discard everything the meter sends until it goes quiet."""
with self._visa.lock():
self._visa.write_raw(_DEVICE_CLEAR)
try:
with self._visa.temporary_timeout(_PROBE_TIMEOUT_MS):
while True:
self._visa.read()
except pyvisa.errors.VisaIOError as e:
# A timeout is the only signal that the meter has gone quiet; anything else is a real fault.
if e.error_code != pyvisa.constants.StatusCode.error_timeout:
raise
def _detect_prompts(self) -> None:
"""Send ``*CLS`` and record whether the meter answers each command line with a prompt.
The manual says prompts are only sent with Echo on (a front-panel-only setting), so
detect them rather than assume.
"""
with self._visa.lock():
self._visa.write("*CLS")
try:
with self._visa.temporary_timeout(_PROBE_TIMEOUT_MS):
prompt = self._read_response("*CLS")
except pyvisa.errors.VisaIOError as e:
if e.error_code != pyvisa.constants.StatusCode.error_timeout:
raise
# A prompt that lands just after the window would otherwise be read as the next *ESR? reply.
self._device_clear()
self._prompts = False
return
self._prompts = True
self._check_prompt("*CLS", prompt)
def _transact(self, command: str) -> str | None:
"""Send one command line, check it executed, and return its response (``None`` for a non-query)."""
with self._visa.lock():
try:
if self._prompts:
return self._transact_prompted(command)
if command.endswith("?"):
self._visa.write(command)
return self._read_response(command)
# Without prompts, chain *ESR? so the reply both confirms completion and carries errors.
line = f"{command}; *ESR?"
self._visa.write(line)
reply = self._read_response(line)
except pyvisa.errors.VisaIOError:
# A reply that arrives after a timeout would otherwise be read as the next command's.
self._device_clear()
raise
try:
esr = int(reply)
except ValueError:
self._device_clear()
raise InstroError(f"Fluke 8808A sent unexpected *ESR? reply {reply!r} for {command!r}") from None
if esr & _ESR_ERROR_MASK:
raise InstroError(f"Fluke 8808A rejected {command!r}: *ESR? = {esr}")
return None
def _transact_prompted(self, command: str) -> str | None:
self._visa.write(command)
response: str | None = None
while True:
line = self._read_line()
if line == _PROMPT_OK or line in _PROMPT_ERRORS:
self._check_prompt(command, line)
return response
if line.upper() != command.upper():
response = line
def _read_response(self, command: str) -> str:
"""Read the response line to ``command``, skipping its echo when Echo is on."""
while True:
line = self._read_line()
if line.upper() != command.upper():
return line
def _read_line(self) -> str:
"""Read the next non-blank line; the meter terminates lines with CR LF."""
while True:
line = self._visa.read().strip()
if line:
return line
@staticmethod
def _check_prompt(command: str, prompt: str) -> None:
if prompt == _PROMPT_OK:
return
reason = _PROMPT_ERRORS.get(prompt, f"unexpected reply {prompt!r}")
raise InstroError(f"Fluke 8808A {reason} for {command!r}")